Faculty of Polytechnics, University of Zenica , Zenica , Bosnia and Herzegovina
LIND-Laboratory for product safety testing , Zenica , Bosnia and Herzegovina
LIND-Laboratory for product safery testing Zenica , Zenica , Bosnia and Herzegovina
LIND-Laboratory for product safety testing Zenica , Zenica , Bosnia and Herzegovina
University of Sarajevo , Sarajevo , Bosnia and Herzegovina
This paper describes the application of international standards, recommendations and guides, as well as other input data, to determine the interval of recalibration of laboratory equipment in the laboratory for testing the product safety testing LIND in Zenica, Bosnia and Herzegovina. An example of a pressure sensor on a 4-axis device for measuring the geometry of furniture, a detailed analysis of the recalibration interval was performed, which had to be shortened from 5 to 3 years.
This is an open access article distributed under the Creative Commons Attribution License which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
The statements, opinions and data contained in the journal are solely those of the individual authors and contributors and not of the publisher and the editor(s). We stay neutral with regard to jurisdictional claims in published maps and institutional affiliations.